Safety standards
- SFDA classification:II
- CE classification:II b
- Type of protection against electric shock: II, with internal power device
- Degree of protection against electric shocks
- Degree of protection against ingress of liquid: IPX1
- Equipment type: Handheld
Mechanical
- Dimension: 58.5mm (w)*123mm (H)*28mm (D)
- Weight :< 200g
Battery
- Type: 3AAA lkaline batteries or
- Lithium ion rechargeable battery (Option)
- Runtime :> 16 hours for normal measurement status
Display
- 2.4” OLED 320*240 pixels
Sound
- Audible alarm, button tone and beep tone
- Supports Pitch Tone and multi-level volume
- Alarm tones meet the requirement of IEC 60601-1-8
Data Storage
- Displaying way: Trend tabular
- Trend interval: 2 seconds to 30 minutes
- Trend parameter: PR, SpO2
- Trend data: Spot-checkmode: ID from 1 To 99,300 groups for each ID
- Continuous monitor mode: 60000 groups can be stored
- Patient Range : Adult, Pediatric and neonatal patients
Standard Digital SpO2
- Range: 0~100%
- Resolution: 1%
- Accuracy: 70% to 100%: ±2%
0% to 69%: unspecified
- Refreshing rate :< 13 seconds
- Pitch Tone: yes
PR
- Range: 25 bum to 250 bpm
- Resolution: 1bpm
- Accuracy: ±2% or +1bpm, witchever is the greater
- Refreshing rate :< 13 seconds
Nellcor SpO2 (Option)
- Range: 0% to 100%
- Resolution: 1%
- Accuracy: 70% to 100%: ±3 %( Neonatal), ±2 %( Adult/Child)
0% to 69%: unspecified
- Refreshing rate: 7 seconds
- Pith Tone: Yes
PR
- Range: 20 bpm to 300 bpm
- Resolution: 1 pum
- Accuracy: 20bpm~250bpm ± 3bpm
251~300bpm unspecified
- Refreshing rate: 7s
ECG
- Lead type: Standard: 3 lead (RA, LA, LL or R, L, F)
- Lead:Ⅰ,Ⅱ,Ⅲ
- Gain: 2.5mm/mV (×0.25), 5mm/mV (×0.5), 10mm/mV (×1)
- Input impedance: ≥5.0MΩ
- Input current: < 0.1uA
- Baseline recovery: Monitor mode: ≤3s
- Electrode offset potential: ±500 mV d.c
- ECG signal input range: -6.0Mv to + 6.0mV
- CMRR: Diagnostic mode : ≥90dB
- Standardizing signal: 1mV ±5%
- Electrode off indicating: Yes
- Work mode: Monitor and Diagnostic mode
- HR range: 10 bpm ~ 300 bpm
- HR Resolution: 1bpm
- HR Accuracy: ±1% or ±1 bpm, whichever is the greater
Standard Parameter:
Standard Digital SpO2, Pulse Rate, Plath bar & SpO2 waveform,
Option:
Nellcor SpO2, 3-lead ECG, infrared receiving adapter, Rechargeable lithium battery (include battery charger & AC-DC adapter)