handheld pulse oximeter BSP-258
 
Safety standards                                            
-         SFDA classification:II                                      
-         CE classification:II b                                        
-         Type of protection against electric shock: II, with internal power device           
-         Degree of protection against electric shocks                   
-         Degree of protection against ingress of liquid: IPX1
-         Equipment type: Handheld                                    
                                                          
Mechanical                                              
-         Dimension: 58.5mm (w)*123mm (H)*28mm (D)                   
-         Weight :< 200g 
                                          
Battery                                                   
-         Type: 3AAA lkaline batteries or                                
-         Lithium ion rechargeable battery (Option)                       
-         Runtime :> 16 hours for normal measurement status              
                                                       
Display                                                 
-         2.4” OLED   320*240 pixels                                
                                                      
Sound                                                  
-         Audible alarm, button tone and beep tone                  
-         Supports Pitch Tone and multi-level volume      
-         Alarm tones meet the requirement of IEC 60601-1-8
                                                         
Data Storage                                 
-         Displaying way: Trend tabular
-         Trend interval: 2 seconds to 30 minutes                        
-         Trend parameter: PR, SpO2                          
-         Trend data: Spot-checkmode: ID from 1 To 99,300 groups for each ID 
-         Continuous monitor mode: 60000 groups can be stored   
-         Patient Range : Adult, Pediatric and neonatal patients
                 
Standard Digital SpO2
-         Range: 0~100%
-         Resolution: 1%
-         Accuracy: 70% to 100%: ±2%
 0% to 69%: unspecified
-         Refreshing rate :< 13 seconds
-         Pitch Tone: yes
PR
-         Range: 25 bum to 250 bpm
-         Resolution: 1bpm
-         Accuracy: ±2% or +1bpm, witchever is the greater
-         Refreshing rate :< 13 seconds
 
Nellcor SpO2 (Option)
-         Range: 0% to 100%
-         Resolution: 1%
-         Accuracy: 70% to 100%: ±3 %( Neonatal), ±2 %( Adult/Child)
0% to 69%: unspecified
-         Refreshing rate: 7 seconds
-         Pith Tone: Yes
 
PR
-         Range: 20 bpm to 300 bpm
-         Resolution: 1 pum
-         Accuracy: 20bpm~250bpm ± 3bpm
                 251~300bpm unspecified
-         Refreshing rate: 7s
 
ECG
-         Lead type: Standard: 3 lead (RA, LA, LL or R, L, F)
-         Lead:Ⅰ,Ⅱ,Ⅲ
-         Gain: 2.5mm/mV (×0.25), 5mm/mV (×0.5), 10mm/mV (×1)
-         Input impedance: ≥5.0MΩ
-         Input current: < 0.1uA
-         Baseline recovery: Monitor mode: ≤3s
-         Electrode offset potential: ±500 mV d.c
-         ECG signal input range: -6.0Mv to + 6.0mV
-         CMRR: Diagnostic mode : ≥90dB
-         Standardizing signal: 1mV ±5%
-         Electrode off indicating: Yes
-         Work mode: Monitor and Diagnostic mode
-         HR range: 10 bpm ~ 300 bpm
-         HR Resolution: 1bpm
-         HR Accuracy: ±1% or ±1 bpm, whichever is the greater
 
Standard Parameter:
Standard Digital SpO2, Pulse Rate, Plath bar & SpO2 waveform,
 
Option:
Nellcor SpO2, 3-lead ECG, infrared receiving adapter, Rechargeable lithium battery (include battery charger & AC-DC adapter)
 
*Specification subject to be changed without notice
 