Main feature:
Abide by the Standard of GB 9706.29-Medical electrical equipment―Part 2: Particular requirements for the safety and essential performance of anaesthsia systems
Be used for closed、semi-closed、semi-open inhalation anaesthesia, controlled or assisted ventilation, monitoring and display of the ventilation parameters.
Applied for adult and child.
Compacted breathing system meets the standard of ISO 8835-2. Good airtight, easy to operation, clearance and sterilization.
Vaporizer meets the standard of ISO 8835-4. Accurate concentration, temperature、flow and pressure compensation technology. Various options of Isoflurane, Enflurane.
Anaesthesia ventilator meets the standard of ISO 8835-5.
Gas driven and electrically controlled
IPPV, MANUAL ventilation
Alarm, indication and protection functions for high pressure、leadoff and apnea
Main technical specification:
●Power : AC 22OV,50Hz
●Gas: O2 Working pressure: 0.28MPa~0.6MPa
Flowmeter
――Adjusting range
O2 flowmeter : 0.1~10L/min
Vaporizer: (Isoflurane, Enflurane. Double or single vaporizer option)
Range of concentration: 0~5%
temperature、flow and pressure compensation technology
●Adjusting range of main ventilation parameters
Frequency: 6/min~60/min
I/E Ratio: 2:1~1:6
Tidal Volume: 0.05~1.5L
●Ventilation parameter monitoring:
Digital display: tidal volume, minute ventilation, frequency, airway pressure
Alarm, indication and protection for;
Power off, low voltage of battery, minute ventilation beyond the limit, pressure beyond the limit, and insufficiency of O2, etc. |